微型黑碳仪AL30
- 产品规格:
- 发货地:广东省深圳市
供应微型黑碳仪AL30 Aethlabs microAeth AL30
产品联系人:黄工

特点 | 黑碳仪AL30 | 黑碳仪MA200 | 黑碳仪MA300 | 黑碳仪MA350 |
测量波长nm | 880 | 880 625 528 470 375 | 880 625 528 470 375 | 880 625 528 470 375 |
测量范围mg BC/m3 | 0-1 | 0-1 | 0-1 | 0-1 |
测量精度μg BC/m3 | ±0.1 | |||
测量分辨率μg BC/m3 | 0.001 | 0.001 | 0.001 | 0.001 |
测量周期s | 1 10 30 60 300 | 1 5 10 30 60 300 | 1 5 10 30 60 300 | 1 5 10 30 60 300 |
流量mL/min | 50 100 150 200 | 50 75 100 125 150 175 | 50 75 100 125 150 175 | 50 75 100 125 150 175 |
重量g | 280 | 420 | 715 | 965 |
该仪器一次充电可连续运行长达24小时。空气样品在聚四氟乙烯(PTFE)过滤介质上收集。过滤介质安装在易于更换的滤条中,便于现场处理。microAeth AL30 可用于测量炉灶、个人暴露以及飞机、火车、自行车、无人机和气象气球。
microAeth / AL30
- Measurement Method
- Real-time Aethalometer method by measuring the rate of change in absorption of transmitted light due to continuous collection of aerosol deposit on filter. Measurement at 880 nm interpreted as concentration of Black Carbon (‘BC’).®
- Measurement Wavelength
- 880 nm
- Timebases (User setting)
- 1, 5, 10, 30, 60, or 300 seconds
- Flow Rates (User setting)
- Internal pump provides 25, 50, 75, 100, 125, 150, 170, 200, 225, or 250 ml/min
- Measurement Range
- 0-1 mg BC/m3, filter life time dependent on concentration and flow rate setting:
avg. 5 μg BC/m3 for 24 hours @ 100 ml/min
avg. 100 μg BC/m3 for 3 hours @ 50 ml/min
avg. 1 mg BC/m3 for 15 minutes @ 50 ml/min
- Measurement Resolution
- 0.001 μg BC/m3
- Limit of Detection
- 0.010 µg BC/m3, 5 minute timebase, 250 ml/min flow rate <0.003 µg BC/m3, 1 hour average from 5 min timebase, 250 ml/min flow rate
- Pump
- Internal micro rotary vane pump
- Flow Control
- Internal mass flowmeter with closed-loop control
- Filter Material / Capacity
- Polytetrafluoroethylene (PTFE) filter strip material with 1 sampling location per filter
- Sampling
- 3 mm spot created on filter strip containing insert of PTFE filter material. PM2.5 size selective inlet available.
- Environmental Sensors
- Temperature, Relative Humidity, Altimeter/Barometer, Accelerometer
- Dimensions and Weight
- Length: 98.5mm (3.878 inches)
Width: 61mm (2.401 inches)
Depth: 24.5mm (0.964 inch)
Weight: 162 grams (5.71 ounces) - Memory
- 8 MB internal flash memory, providing data storage up to:
30 days operating on 5 minute timebase
7 days operating on 1 minute timebase
- Wireless Communications / Data Output
- Data Stored in local memory.
Direct local data download over WiFi interface.
WiFi transmission to cloud data management system. Real-time and bulk data transfer when network is available 802.11 b/g/n Wi-Fi - Connections
- USB-C (Power)
3.3V TTL Serial (Flow Calibration Communication)
Aerosol sample inlet and outlet ports - Total Run Time (Single battery charge)
- Up to 24 hours @ 5 minute timebase, 50 ml/min flow rate. Run time may vary due to settings and PM concentrations.
- Battery
- Internal 3.7V 2200 mAh (8.14 Wh) 1 cell rechargeable lithium-ion battery.
- Charging Time
- Approximately 2.2 hours to full charge (using USB-C AC adapter, instrument turned off).
- Power Supply Adapter
- Input: 100~240 VAC 50/60 Hz 0.4 A
Output: 5VDC / 2A, USB-C - Operation Environment
- 5 ~ 40 °C operating, non-condensing.
- Included
microAeth AL30
1 Pack of 30 Filter Strips
1 USB-C power cable
1 territory-specific USB-C AC wall adapter
1 one meter sampling hose with swivel tube connector
1 Lapel clip for sampling hose
Manual available for download via AethLabs website
m.octsensor.b2b168.com